Print

Lab 9 - Solubility Product Constants

Purpose

To measure the solubility product constant (Ksp) of copper (II) iodate, Cu(IO3)2.

Goals

Introduction

In previous introductory chemistry courses, you learned some basic solubility rules that are useful in determining if an ionic solid will dissolve in water. Solids that dissolve completely, such as NaCl and NH4NO3, were referred to as "soluble" and others that did not dissolve completely, such as AgCl and BaSO4, were referred to as "insoluble." In fact, very few ionic solids are completely insoluble, meaning that they will not form any ions when placed in aqueous solution. Most solids that are commonly referred to as "insoluble" are actually slightly soluble and will produce an equilibrium between undissolved solid and ions in solution. For example, when copper (II) iodate (Cu(IO3)2) is placed in water, the following equilibrium is established.
( 1 )
Cu(IO3)2(s) equilibrium arrow Cu2+(aq) + 2 IO3(aq)
 
The equilibrium constant associated with this reaction is called the solubility product constant and is given the symbol Ksp.
( 2 )
Ksp = [Cu2+][IO3 ]2 
It is important to emphasize that the equilibrium in equation 1
Cu(IO3)2(s) equilibrium arrow Cu2+(aq) + 2 IO3(aq)
 
is only true if some solid is present. If the solid completely dissolves in solution, then the product of the ions as shown in equation 2
Ksp = [Cu2+][IO3 ]2 
is not equal to the Ksp. However, as long as some solid is in contact with solution, the solution will become saturated with the ions according to equation 1
Cu(IO3)2(s) equilibrium arrow Cu2+(aq) + 2 IO3(aq)
 
.
The molar solubility of a solid is the maximum number of moles of the solid that will dissolve in one liter of solution. Molar solubility is measured in moles/liter and has units of molarity (M). Molar solubility can be determined by measuring the concentration of the ions formed in a solution saturated with the solid. For example, if the molarity of Cu2+ in a solution saturated with Cu(IO3)2 can be determined, the stoichiometry in equation 1
Cu(IO3)2(s) equilibrium arrow Cu2+(aq) + 2 IO3(aq)
 
indicates that this is also equal to the molarity of Cu(IO3)2 that dissolved in solution, or its molar solubility. Alternatively, if the molarity of IO3 in a solution saturated with Cu(IO3)2 can be determined, the stoichiometry in equation 1
Cu(IO3)2(s) equilibrium arrow Cu2+(aq) + 2 IO3(aq)
 
indicates that the molar solubility of Cu(IO3)2 in water is one half the [IO3] in solution.
In a saturated solution, if all the ions came from the solid, then the ratio of cations to anions is known. In the case of Cu(IO3)2, it is known that 2[Cu2+] = [IO3]. Thus, if the Cu2+ concentration can be measured, the IO3 concentration can be calculated and vice versa. These concentrations can then be entered into the Ksp expression (equation 2
Ksp = [Cu2+][IO3 ]2 
) to solve for the solubility product constant of Cu(IO3)2.
The preceding discussion referred to dissolving the sparingly soluble salt in pure water. However, if some of the ions that are to be produced by the solid are present in solution from another source, Le Châtelier's Principle predicts that the equilibrium in equation 1
Cu(IO3)2(s) equilibrium arrow Cu2+(aq) + 2 IO3(aq)
 
will shift to the left and less solid will dissolve. This would result in a lower molar solubility. This is referred to as the common ion effect. In the example of Cu(IO3)2, the presence of either Cu2+ or IO3 in solution should result in a lower molar solubility than in pure water.
Earlier this semester in the Solutions and Spectroscopy lab, you prepared a calibration curve for Cu2+ over the range of approximately 0.1 M to 0.4 M. As you may recall, Cu2+ in solution has a pale blue color and you were able to measure its absorbance at 620 nm. For this experiment, it is necessary to prepare a calibration curve for Cu2+ over the range of 5 × 10–4 M to 0.01 M. At these concentrations, the absorbance of Cu2+ is too low to detect directly by spectroscopy. To improve the absorbance of the Cu2+ solutions, aqueous ammonia (NH3) will be added to each solution to produce the complex ion Cu(NH3)42+ according to the following reaction.
( 3 )
Cu2+(aq) + 4 NH3(aq) equilibrium arrow Cu(NH3)42+(aq)
 
Cu(NH3)42+ in solution has a dark blue color and its absorbance at 600 nm can be measured over the range necessary for this experiment. The equilibrium constant associated with equation 3
Cu2+(aq) + 4 NH3(aq) equilibrium arrow Cu(NH3)42+(aq)
 
is called the formation constant and is given the symbol Kf. The Kf for Cu(NH3)42+ is 4.8 × 1012 which indicates that the forward direction of this reaction is greatly favored over the reverse. The result is that, with a large excess of NH3 present, almost all of the Cu2+ ions are present as the complex ion, Cu(NH3)42+. Thus we will make the assumption that the [Cu(NH3)42+] is equal to the [Cu2+] in the solution before the addition of NH3.
In this experiment, you will prepare a calibration curve for Cu(NH3)42+ from four standard Cu(NO3)2 solutions. Then you will calculate the solubility product constant for Cu(IO3)2 from measurements of the [Cu2+] in five solutions saturated with Cu(IO3)2. You will also calculate the molar solubility of copper (II) iodate in pure water and in solutions containing Cu2+ and IO3 and compare your results to predictions of the common ion effect.

Equipment

  • 15 13 × 100 mm test tubes
  • 5 stoppers
  • 1 test tube rack
  • 1 spatula
  • 1 glass stir rod
  • 1 MicroLab spectrophotometer
  • 1 MicroLab spectrophotometer instruction sheet
  • 6 vials
  • 1 250 mL waste beaker
  • 1 deionized water squirt bottle
  • 1 box of Kimwipes

Reagents

  • ~0.25 g solid Cu(IO3)2
  • ~5 mL 5.0 × 10-4 M Cu(NO3)2
  • ~10 mL 1.0 × 10-3 M Cu(NO3)2
  • ~10 mL 5.0 × 10-3 M Cu(NO3)2
  • ~5 mL 1.0 × 10-2 M Cu(NO3)2
  • ~5 mL 5.0 × 10-3 M KIO3
  • ~5 mL 1.0 × 10-2 M KIO3
  • ~5 mL 3 M NH3
  • deionized water

Safety

Cu(NO3)2 is listed as an oxidizer and corrosive. KIO3 and Cu(IO3)2 are listed as strong oxidizing agents. As with all chemicals in the lab, if you come in contact with a solid, you should gently brush off the affected area with a paper towel and then flush the area with water. If you come in contact with a solution, you should flush the affected area with water.

Waste Disposal

All solutions generated this week must be placed in the waste bottle in the lab. Designate a "waste" beaker and set it aside for use during your lab. You can put the small samples of solution you will make into this beaker and empty it into the waste bottle at the end of class, instead of going back and forth to the waste bottle. Always remember not to overfill the waste bottle. If your waste bottle is full, please alert your lab instructor.

Prior to Class

Please read the following sections of the Introductory Material:

Lab Procedure

Please print the worksheet for this lab. You will need this sheet to record your data.

PDF file

1
A spectrophotometer will be set up in your work area. Make sure it is plugged in and allow it to warm up.
2
While the spectrophotometer is warming up, label five clean, dry test tubes A - E and set up a beaker for waste.
3
Fill each of the five test tubes to a depth of 4 fingers or approximately 3/4 full with the following solutions.
4
Using the spatula provided, add a small amount of solid Cu(IO3)2 to each of the five test tubes. A "small amount" means just a little bit on the smaller end of the spatula. If the solid dissolves entirely, you can always add some more solid to the test tube.
5
Stopper each of the test tubes. While holding the stopper in place, invert the tubes and agitate the contents for several minutes.
6
Allow the test tubes to equilibrate for a minimum of 10 minutes. Agitate the solutions periodically while you work on your standard solutions.
7
Label four additional clean, dry test tubes 1 - 4. These will be the standards for the calibration curve.
8
Fill each of the four test tubes to a depth of 3 fingers or approximately 1/2 full with the following solutions. Try to fill each test tube to the same level.
9
Add 10 drops of 3 M NH3 to each of the four test tubes (1 - 4).
10
With the glass stir rod provided, gently mix each solution so the dark blue color is uniform throughout the test tube. Be sure to rinse and dry the stir rod between solutions to avoid cross-contamination and errors from dilution.
11
Following the detailed instructions for the spectrophotometer provided in lab, take a background spectrum.
12
Take a blank spectrum with deionized water. To condition your vial, carefully pour some deionized water into a vial and pour it out to waste then refill with deionized water.
13
Condition a vial using standard solution 1 and take an absorbance spectrum. Identify the wavelength of maximum absorbance near 600 nm, and record it. Record the absorbance at this wavelength in Table A. Absorbance values are reported to the 0.001. For all remaining absorbance measurements in this experiment, be sure to use the same wavelength you have identified as the maximum. Retain this sample in the vial until you have completed your calibration plot. Students often choose to label a sheet of paper with positions 1, 2, 3, and 4, placing each vial on the appropriate position.
14
Repeat step 13 for each of the remaining standard solutions and record their absorbances in Table A.
15
The MicroLab software will plot the absorbance of the Cu2+ solutions that you made as a function of their concentrations. The trendline and R2 value are displayed. If your plot is linear with an R2 value of 0.9 or greater, continue the experiment. If your R2 value is low, consult with your lab instructor. Record these in Table A. Do not close the MicroLab file, as this calibration curve will be used to determine Cu2+ equilibrium concentrations.
16
Once you are satisfied with your calibration curve, empty the four standard solution vials into your waste beaker and rinse each one with deionized water.
17
Remove the stoppers from the five test tubes labeled A - E that you prepared in steps 3 - 6. Place them in a centrifuge adding one tube filled 3/4 full with deionized water for balance. Centrifuge the tubes for 2 full minutes. Allow the centrifuge to stop on its own so as not to agitate the solid present.
18
Label five additional clean, dry test tubes A - E.
19
Decant each of the five solutions from the centrifuge into the new test tube of the same label. Fill each to a depth of 3 fingers or 1/2 full as you did for the standard solutions in step 8. Again, try to fill them to approximately the same level. When doing this, be careful not to transfer any solid! Do not start to pour and stop because this will stir up the solid and cause you to have to re-centrifuge the solution. If you over-decant, you may pour some back into the tube containing solid.
20
Add 10 drops of 3 M NH3 to each of the five test tubes (A - E).
21
With the glass stir rod provided, gently mix each solution so the dark blue color is uniform throughout the test tube. Be sure to rinse and dry the stir rod between solutions to avoid cross-contamination and errors from dilution.
22
Condition a vial using solution A, refill the vial, measure its absorbance and record this value in Table B. Be sure to use the same wavelength used in the calibration curve. Do not discard the solution until you have completed your measurements of solutions A - E.
23
Repeat step 22 for each of the remaining solutions and record their absorbances in Table B.
24
Once you are satisfied with your results, empty the five vials into your waste beaker and rinse each one with deionized water.
25
When you are finished taking measurements, close the MicroLab software and unplug your spectrophotometer.
26
Collect all your copper solution waste and place it in the waste bottle in the lab, making sure not to overfill it. Rinse and dry all your glassware with water and return it to the set-up area where you found it.
27
Remember to show your TA your calibration curve and calculated Ksp values. Your TA will manually grade the results and enter your score into WebAssign.
28
Before leaving, enter your results in the in-lab assignment. If all results are scored as correct, log out. If not all results are correct, try to find the error or consult with your lab instructor. When all results are correct, note them and log out of WebAssign. The in-lab assignment must be completed by the end of the lab period. If additional time is required, please consult with your lab instructor.