Lab 10 - Solubility Product Constants
Purpose
To measure the solubility product constant (Ksp) of copper (II) iodate, Cu(IO3)2.Goals
-
1To measure the molar solubility of a sparingly soluble salt in water.
-
2To prepare a calibration curve based on complex ion formation for absorbance enhancement.
-
3To calculate the solubility product constant (Ksp) of a sparingly soluble salt from its molar solubility.
-
4To confirm the common ion effect on the molar solubility of a sparingly soluble salt.
Introduction
In previous introductory chemistry courses, you learned some basic solubility rules that are useful in determining if an ionic solid will dissolve in water. Solids that dissolve completely, such as NaCl and NH4NO3, were referred to as "soluble" and others that did not dissolve completely, such as AgCl and BaSO4, were referred to as "insoluble". In fact, very few ionic solids are completely insoluble, meaning that they will not form any ions when placed in aqueous solution. Most solids that are commonly referred to as "insoluble" are actually slightly soluble and will produce an equilibrium between undissolved solid and ions in solution. For example, when copper (II) iodate (Cu(IO3)2) is placed in water, the following equilibrium is established.( 1 )
Cu(IO3)2(s) Cu2+(aq) + 2 IO3-(aq)
( 2 )
Ksp = [Cu2+][IO3− ]2
( 3 )
Cu2+(aq) + 4 NH3(aq) Cu(NH3)42+(aq)
Equipment
-
1513 x 100 mm test tubes
-
5stoppers
-
1test tube rack
-
1spatula
-
1glass stir rod
-
1MicroLab Spectrophotometer
-
1MicroLab Spectrophotometer Instruction Sheet
-
6vials
-
1250 mL waste beaker
-
1deionized water squirt bottle
-
1box of Kimwipes
Reagents
-
~0.25 gsolid Cu(IO3)2
-
~5 mL5.0 x 10-4 M Cu(NO3)2
-
~10 mL1.0 x 10-3 M Cu(NO3)2
-
~10 mL5.0 x 10-3 M Cu(NO3)2
-
~5 mL1.0 x 10-2 M Cu(NO3)2
-
~5 mL5.0 x 10-3 M KIO3
-
~5 mL1.0 x 10-2 M KIO3
-
~5 mL3 M NH3
- deionized water
Safety
Cu(NO3)2 is listed as an oxidizer and corrosive. KIO3 and Cu(IO3)2 are listed as strong oxidizing agents. NH3 is corrosive. These chemicals can attack the skin and cause permanent damage to the eyes. If one of these solutions splashes into your eyes, use the eyewash immediately. Hold your eyes open and flush with water for at least 15 minutes. If contact with skin or clothing occurs, rinse the affected area with water for at least 15 minutes. Have your lab partner notify your teaching assistant about the spill and your exposure. The NH3 solution also has irritating vapors. It should be used in a fume hood to avoid inhaling the vapors. As with all chemicals in the lab, if you come in contact with a solid, you should gently brush off the affected area with a paper towel and then rinse the area with water. If you come in contact with a solution, you should rinse the affected area with water. Students will have access to gloves due to the use of the concentrated ammonia solution during the lab period.Waste Disposal
All solutions generated this week must be placed in the waste bottle in the lab. Designate a "waste" beaker and set it aside for use during your lab. You can put the small samples of solution you will make into this beaker and empty it into the waste bottle at the end of class, instead of going back and forth to the waste bottle. Always remember not to overfill the waste bottle. If your lab section's waste bottle is full, please alert your teaching assistant.Prior to Class
Please read the following sections of the Introductory Material: Please complete your WebAssign prelab assignment. Check your WebAssign account for due dates. Students who do not complete the WebAssign prelab assignment are required to print and hand in the prelab worksheet.Lab Procedure
Please print the worksheet for this lab. You will need this sheet to record your data.1
A spectrophotometer will be set up in your work area. Make sure it is plugged in and allow it to warm up.
2
While the spectrophotometer is warming up, label five clean, dry test tubes A - E and set up a beaker for waste.
3
Fill each of the five test tubes to 3/4 full with the following solutions.
4
Using the spatula provided, add a small amount of solid Cu(IO3)2 to each of the five test tubes. A "small amount" means just a little bit on the smaller end of the spatula. If the solid dissolves entirely, you can always add some more solid to the test tube.
5
Stopper each of the test tubes. While holding the stopper in place, invert the tubes and agitate the contents for several minutes.
6
Allow the test tubes to equilibrate for a minimum of 10 minutes. Agitate the solutions periodically while you work on your standard solutions.
7
Label four additional clean, dry test tubes 1 - 4. These will be the standards for the calibration curve.
8
Fill each of the four test tubes to 2/3 full with the following solutions. Try to fill each test tube to the same level.
9
Add 12 drops of 3 M NH3 to each of the four test tubes (1 - 4).
10
With the glass stir rod provided, gently mix each solution so the dark blue color is uniform throughout the test tube. Be sure to rinse and dry the stir rod between solutions to avoid cross-contamination and errors from dilution.
11
Following the detailed instructions for the spectrophotometer provided in lab, take a background spectrum.
12
Take a blank spectrum with deionized water. To condition your vial, carefully pour some deionized water into a vial and pour it out to waste then refill with deionized water.
13
Condition a vial using standard solution 1 and take an absorbance spectrum. Identify the wavelength of maximum absorbance near 600 nm. Record the wavelength and absorbance at this wavelength in Data Table A. Absorbance values are reported to the 0.0001. For all remaining absorbance measurements in this experiment, be sure to use the same wavelength you have identified as the maximum. Retain this sample in the vial until you have completed your calibration plot. Students often choose to label a sheet of paper with positions 1, 2, 3, and 4, placing each vial on the appropriate position.
14
Repeat step 13 for each of the remaining standard solutions and record their absorbances in Data Table A.
Table A: Absorbances of Standard Cu2+ Solutions
15
The MicroLab software will plot the absorbance of the Cu2+ solutions that you made as a function of their concentrations. The trendline and R2 value are displayed. If your plot is linear with an R2 value of 0.9 or greater, continue the experiment. If your R2 value is low, consult with your teaching assistant. Record these in Data Table A. Do not close the MicroLab file, as this calibration curve will be used to determine Cu2+ equilibrium concentrations.
16
Once you are satisfied with your calibration curve, empty the four standard solution vials into your waste beaker and rinse each one with deionized water.
17
Remove the stoppers from the five test tubes labeled A - E that you prepared in steps 3 - 6. Place them in a centrifuge along with one additional tube filled 3/4 full with deionized water as a counter balance. Centrifuge the tubes for 2 full minutes. Allow the centrifuge to stop on its own so as not to agitate the solid present.
18
Label five additional clean, dry test tubes A - E.
19
Decant each of the five solutions from the centrifuge into the new test tube of the same label. Fill each to 2/3 full as you did for the standard solutions in step 8. Again, try to fill them to approximately the same level. When doing this, be careful not to transfer any solid! Do not start to pour and stop because this will stir up the solid and cause you to have to re-centrifuge the solution. If you over-decant, you may pour some back into the tube containing solid.
20
Add 12 drops of 3 M NH3 to each of the five test tubes (A - E).
21
With the glass stir rod provided, gently mix each solution so the dark blue color is uniform throughout the test tube. Be sure to rinse and dry the stir rod between solutions to avoid cross-contamination and errors from dilution.
22
Condition a vial using solution A, refill the vial, and measure its absorbance. Remember that these next samples are not part of your calibration curve and should be read as unknown concentrations. Record this value in Data Table B. Be sure to use the same wavelength used in the calibration curve. Do not discard the solution until you have completed your measurements of solutions A - E.
23
Repeat step 22 for each of the remaining solutions and record their absorbances in Data Table B.
Table B: Absorbances of Solutions Saturated with Cu(IO3)2
24
Once you are satisfied with your results, empty the five vials into your waste beaker and rinse each one with deionized water.
25
When you are finished taking measurements, close the MicroLab software and unplug your spectrophotometer.
26
Collect all your copper solution waste and place it in the waste bottle in the lab, making sure not to overfill it. Rinse and dry all your glassware with water and return it to the set-up area where you found it.
Question 1: Using the trendline from your calibration curve, calculate the [Cu2+] in each of the solutions A - E. Show one representative calculation neatly. Enter these results in Data Table B.
Question 2: Calculate the [IO3-] in each of the solutions A - E. Remember that for solutions A and B where IO3- was originally present, you must account for the initial amount plus any ions that were formed due to the solubility of the solid. Show one representative calculation neatly. Enter these results in Data Table B.
Question 3: Calculate the Ksp for each of the solutions A - E. Show one representative calculation neatly. Enter these results in Data Table C.
Table C: Calculated Ksp and Solubilities of Cu(IO3)2
Question 4a: Calculate the average of your five Ksp values. Enter the result in Data Table C.
Question 4b: The literature value for the Ksp of Cu(IO3)2 is 6.85 x 10-8. How does this compare to your average experimental values? What is the percent error compared to your average?
-
% error =
x 100experimental - theoretical theoretical
Question 5: Calculate the solubility of Cu(IO3)2 in each of the five solutions. Remember that for solutions D and E where Cu2+ was originally present, you must subtract the initial amount from the equilibrium amount to get the amount formed due to the solubility of the solid. Show one representative calculation neatly. Enter these results in Data Table C.
Question 6: Looking at your solubility results for solutions A and B as compared to C, did the solubility increase, decrease or stay about the same in the solutions that originally contained IO3- ions? Does this confirm the common ion effect? If not, can you explain why it does not?
Question 7: Looking at your solubility results for solutions D and E as compared to C, did the solubility increase, decrease or stay about the same in the solutions that originally contained Cu2+ ions? Does this confirm the common ion effect? If not, can you explain why it does not?
28
Remember to show your TA your calibration curve and calculated Ksp values. Your TA will manually grade the results and enter your score into WebAssign.
29
Before leaving, enter your results in the InLab assignment. If all results are scored as correct, log out. If not all results are correct, try to find the error or consult with your teaching assistant. When all results are correct, note them and log out of WebAssign. The InLab assignment must be completed by the end of the lab period. If additional time is required, please consult with your teaching assistant.